Appendix D. Output Tables
TABLE D-4. Variables of the RawDataTable
Name Units Statistic
When Used in
Calculation
When Included in Table
T_air(8) °C Smp last sample N_AirTemps > 7
pump_control Smp last sample SaveAll_diagnostics = True
pump_speed Hz Smp last sample SaveAll_diagnostics = True
pump_tmpr °C Smp last sample SaveAll_diagnostics = True
pump_heat Smp last sample SaveAll_diagnostics = True
pump_fan Smp last sample SaveAll_diagnostics = True
valve_tmpr °C Smp last sample SaveAll_diagnostics = True
valve_heat Smp last sample SaveAll_diagnostics = True
valve_fan Smp last sample SaveAll_diagnostics = True
intake_heat Smp last sample SaveAll_diagnostics = True
STARTsequence Smp last sample SaveAll_diagnostics = True
STOPsequence Smp last sample SaveAll_diagnostics = True
seq_ACTIVE Smp last sample SaveAll_diagnostics = True
valve_mode Smp last sample SaveAll_diagnostics = True
ZeroSpanMode Smp last sample SaveAll_diagnostics = True
DO_zero Smp last sample SaveAll_diagnostics = True
DO_CO2_span Smp last sample SaveAll_diagnostics = True
DO_H2O_span Smp last sample SaveAll_diagnostics = True
interval_counter Smp last sample SaveAll_diagnostics = True
end_omit Smp last sample SaveAll_diagnostics = True
sequence_index Smp last sample SaveAll_diagnostics = True
NewSite Smp last sample SaveAll_diagnostics = True
SiteOutput Smp last sample SaveAll_diagnostics = True
SiteAvgDisable Smp last sample SaveAll_diagnostics = True
TimeInfo
The TimeInfo table stores all of the timing parameters for the automatic valve
switching sequence. A record is stored whenever the profile sequence starts
(when the program first starts up, and when the sequence is manually stopped
and restarted). This table can generally be ignored, but it provides useful
historical information about the valve sequence timing. It may be helpful for
troubleshooting sequence timing issues.
The first values stored are the system configuration variables that control the
sequence timing: LEVELS_USED, MEASURE_TANKS,
D-9