Appendix J. AP200 Pump Replacement
FIGURE J-4. Location of pump connector in AP200 pump electronics
5. Gently lift the pump assembly from foam, leaving the tubing attached.
6. Turn assembly over and remove the two self-tapping #6 screws that
attach the pump to the metal box, as shown in FIGURE J-5. If these
screws become lost or damaged, replace them with pn 13535.
FIGURE J-5. Self-tapping screws attaching pump to metal box
7. Cut the blue (inlet) and red (outlet) tubing on each side of the pump
behind the barbed connector as shown in FIGURE J-6.