FIGURE 7-2. Order of replacement for orifice, filter, and rain diverter
Thread the rain diverter onto the filter and slide the insulation back down the
tube to the filter. Press the rain diverter back through its opening in the intake
assembly. See FIGURE 7-3.
Take care not to loosen the connections between the rain
diverter, filter, orifice, and tube when snapping the rain diverter
back into its mounting hole.
Restart the pump and check the sample flow. If replacing the filter has not
restored the flow, contact Campbell Scientific.
FIGURE 7-3. Correct reassembly after filter replacement
NOTE
49