Section 8. Operation
313
Figure 98: Input error calculation
Input Error Examples: Type T Thermocouple @ 45°C
These examples demonstrate that in the environmental temperature range, input-
offset error is much greater than input-gain error because a small input range is
used.
Conditions:
CR3000moduletemperature,‐25to50°C
Temperature=45°C
Referencetemperature=25°C
DeltaT(temperaturedifference)=20°C
Thermocoupleoutputmultiplierat45°C=42.4µV°C
‐1
Thermocoupleoutput=20°C*42.4µV°C
‐1
=830.7µV
Inputrange=±20mV