Section 8. Operation
304
Offset = 3 x Basic Resolution + 5.0 µV if the measurement is of a single-
ended input channel
Thefollowingtablelistsbasicresolutionvalues.
Table 65. Analog Input-Voltage Range and Basic
Resolution
Range (mV) Basic Resolution (µV)
±5000
±1000
±200
±50
±20
167
33.3
6.67
1.67
0.67
Assumptions that support the ratiometric-accuracy specification include:
Excitation voltages less than 500 mV and excitation currents less than 500
µA are reversed during the excitation phase of the measurement.
Effects due to the following are not included in the specification:
o Bridge-resistor errors
o Sensor noise
o Measurement noise
The ratiometric-accuracy specification is applied to a three-wire half-bridge
measurement that uses the BrHalf() instruction as follows:
TherelationshipdefiningtheBrHalf()instructionisX=V1/Vx,whereV1
isthevoltagemeasurementandVxistheexcitationvoltage.The
estimatedaccuracyofXisdesignatedas∆X,where∆X=∆V1/Vx.∆V1is
derivedusingth
efollowingmethod.
The ratiometric-accuracy specification is applied to a four-wire full-bridge
measurement that uses the BrFull() instruction as follows:
TherelationshipdefiningtheBrFull()instructionisX=1000*V1/Vx,
whereV1isthevoltagemeasurementandVxistheexcitationvoltage.
ResultXisexpressedasmV/V.EstimatedaccuracyofXis∆X,where∆X
=1000*V1/Vx.∆V1isderiv
edusingthefollowingmethod.
V1isderivedusingtheratiometricaccuracyequation.Thederivation
isillustratedinthisexample,whichissupportedbytheassumptionthat
themeasurementisdifferentialwithinputreversal,datalogger
temperatureisbetweento40°C,analoginputrangeis±200mV,V1=
11
0mV,andexcitationisreversedduringtheexcitationphaseofthe
measurement.Theeffecteachassumptionhasonthemagnitudeof
V1inthisexampleisnotedinthefollowingfigure.
In the case of the Resistance() instruction, sensor resistance is determined from
Vs/Ix, where excitation current Ix is measured across a 1000 , ±0.005% @ 25
˚C, 2 ppm/˚C TCR internal resistor.