Section 8. Operation
278
Table 51. CRBasic Parameters Varying Measurement Sequence and
Timing
CRBasic Parameter Description
MeasOfs
Correct ground offset on single-ended measurements.
RevDiff
Reverse high and low differential inputs.
SettlingTime
Sensor input settling time.
Integ
Duration of input signal integration.
RevEx
Reverse polarity of excitation voltage.
8.1.2.4 Measurement Accuracy
CR1000 analog-measurement error is calculated as
Error=GainError(%)+OffsetError
Gain error is expressed as ±% of reading and is a function of CR1000
temperature. Between 0°C and 40°C, gain error is ±0.06% of input voltage. This
gain error assumes factory recalibration every two years.
Offset error depends on measurement type and input range. For differential
measurements with input reversal,
OffsetError=1.5BasicResolution+1.0µV
where
BasicResolutionisthepublishedresolutionoftheprogrammedinput
voltagerange(seeCR1000Specifications
).
Figure Voltage Measurement Accuracy (0° to 40°C)
(p. 279) illustrates that as
magnitude of input voltage decreases, measurement error decreases.
Note The accuracy specification includes only the CR1000 contribution to
measurement error. It does not include the error of sensors.
For example, assume the following (see CR1000 Specifications ):
Input Voltage: 2500 mV
Programmed Input Voltage Range: ±2500 mV (mV2500)
Programmed Measurement Instruction: VoltDiff()
Input Measurement Reversal: True
CR1000 Temperature: Between 0°C and 40°C
Accuracy of the measurement is calculated as follows:
Error=GainError+OffsetError,