Table 8. Serial String Format (Continued)
5 - 7 ON-TIME CNT Power-On Time (In Hex) in 10-Minute Increments 00 - 01FFFF
8 - 9 COLD START CNT Cold Start Power-On Count 00 - FFFF
10-11 ICE CNT Ice Events 00 - FFFF
12 - FAIL CNT Total Failures Encountered. This number is incremented each
time the ice detector transitions from OK to fail.
00 - FF
13 - FAIL DTL 1 7 - 4 MSO Frequency Fail Count 0 - F
3 - 0 Heater Fail Count 0 - F
14 - FAIL DTL 2 7 - 4 Not Used Not Used
3 - 0 Not Used Not Used
15 - LAST ERR 1 See ERRSTAT1 Above
16 - LAST ERR 2 See ERRSTAT2 Above
17 - 2ND LAST ERR 1 See ERRSTAT1 Above
18 - 2ND LAST ERR 2 See ERRSTAT2 Above
19 - PERM ERR 1 See ERRSTAT1 Above
20 - PERM ERR 2 See ERRSTAT2 Above
21 - Software Version 7 - 0 Software Version per VDD/SC1 0 - FF
22 - Correlation Count 7-0 0.01" ice accretion increments since power-on 0 - FF
23 - CHECKSUM Summation (1-byte wide) of bytes 0 - 22 0 - FF
10 Electrostatic Discharge (ESD) Consideration
The freezing rain sensor internal components are ESD sensitive, class 1, so proper ESD precautions must be observed
(wrist straps, conductive surfaces) when handling.