Pagina 1 - INSTRUCTION MANUAL

INSTRUCTION MANUAL Model 109SS Temperature Probe Revision: 3/14 Copyright © 1983- 2014 Campbell Scientific, Inc.

Pagina 2

Model 109SS Temperature Probe 8. If LoggerNet, PC400, RTDAQ, or PC200W is running on your PC, and the PC to datalogger connection is active, you can

Pagina 3 - Limited Warranty

Model 109SS Temperature Probe Maximum Steinhart-Hart Linearization Error: 0.02 °C at –40 °C FIGURE 6-1. Worst-case probe and measurement er

Pagina 4 - Assistance

Model 109SS Temperature Probe Stainless-Steel Sheath Diameter: 0.16 cm (0.063 in) Length: 5.84 cm (2.3 in) Overmolded Joint Diameter: 1.02

Pagina 5 - Table of Contents

Model 109SS Temperature Probe Short Cut cannot edit programs after they are imported and edited in CRBasic Editor. A Short Cut tutorial is available i

Pagina 6 - Figures

Model 109SS Temperature Probe 1: AC Half Bridge (P5) 1: 1 Reps 2: 25 2500 mV 60 Hz Rejection Range 3: 1 SE Channel 4: 1

Pagina 7 - 4. Quickstart

Model 109SS Temperature Probe 7.3 Water Temperature Installation 109SS probes can be submerged to 45 m (150 ft) or 63 psi. The 109SS is not weighted

Pagina 8

Model 109SS Temperature Probe The ratio of measured voltage (Vs) to excitation voltage (Vx) is related to thermistor resistance (Rs) and the 24.9 kΩ f

Pagina 9

Model 109SS Temperature Probe where: Tk = temperature in Kelvin X = 0.001 • ln(Rs) C0 = A = 0.001129 C1 = B • 1E3 = 0.234108 C2 = 0 C3 = C • 1E9 = 87.

Pagina 10 - 6. Specifications

Model 109SS Temperature Probe Therm109(T109_C,1,1,1,20000,_60Hz,1.0,0.0) Integration options and the settling time parameter are not available for CR2

Pagina 11 - Temperature Degrees C

Model 109SS Temperature Probe 9.2 Maintenance The 109SS probe requires minimal maintenance. Periodically check cabling for signs of damage and possi

Pagina 13 - 7.2.2 Edlog

Model 109SS Temperature Probe 14

Pagina 14

Appendix A. Importing Short Cut Code This tutorial shows: • How to import a Short Cut program into a program editor for additional refinement. • H

Pagina 15 - 8. Operation

Appendix A. Importing Short Cut Code 6. Import wiring information to the program by opening the associated .DEF file. Copy and paste the section be

Pagina 16

Appendix B. Example Programs B.1 Example CRBasic Programs B.1.1 Example 1 — Sample program for CR200(X) series dataloggers 'Program measures

Pagina 17 - 8.4 Long Cable Lengths

Appendix B. Example Programs '*AG = Analog Ground (represented by ground symbol on CR1000 wiring panel 'Declare the variables for the temp

Pagina 18 - 9.1 Troubleshooting

Appendix B. Example Programs 'Call the data table CallTable AvgTemp NextScan EndProg B.2 Example Edlog Program This example can be u

Pagina 19 - 9.3 Calibration

Appendix B. Example Programs 7: Polynomial (P55) 1: 1 Reps 2: 6 X Loc [ Scal_lnRt ] 3: 7 F(X) Loc [ 1_Tk ] 4: .001129

Pagina 20

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature TABLE C-1. Voltage Ratio, Resistance, and Temperature1 Actual Tempe

Pagina 21 - A.1.1 CRBasic Datalogger

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature 3 28054.4 3.02 3.00 4 26687.5 4.02 4.00 5 25395 5.02 5.00 6 24172.5 6

Pagina 22 - A.1.2 Edlog

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature 59 2578.6 59.03 59.00 60 2487.1 60.03 60.00 61 2399.4 61.03 61.00 62

Pagina 23

Limited Warranty “Products manufactured by CSI are warranted by CSI to be free from defects in materials and workmanship under normal use and service

Pagina 24

Appendix C. Conversion of Thermistor Resistance or Voltage Ratio to Temperature C-4

Pagina 26

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbe

Pagina 27 - Temperature

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Pagina 28

Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to spec

Pagina 29

Table of Contents B. Example Programs ... B-1 B.1 Example CRBasic Programs ...

Pagina 30

Model 109SS Temperature Probe 1. Introduction The 109SS Temperature Probe uses a thermistor to measure temperature in soil and water. It is compati

Pagina 31

Model 109SS Temperature Probe 3. When Short Cut opens, select New Program. 4. Select Datalogger Model and Scan Interval (default of 5 or 10 second

Pagina 32

Model 109SS Temperature Probe 5. Under the Available Sensors and Devices list, select the Sensors | Temperature folder. Select 109 Temperature Probe