Model 109SS Temperature Probe
Maximum
Steinhart-Hart
Linearization Error: 0.02 °C at 40 °C
FIGURE 6-1. Worst-case probe and measurement errors
FIGURE 6-2. Steinhart-Hart linearization error
1
The overall probe accuracy is a combination of the thermistor interchangeability specification and the
accuracy of the bridge resistor. The Steinhart-Hart equation used in CRBasic instruction Therm109()
(CRBasic dataloggers) has negligible error. The major error component is the interchangeability
specification of the thermistor. The bridge resistor has a 0.1% tolerance with a 10 ppm temperature
coefficient.
-0.005
0
0.005
0.01
0.015
0.02
0.025
0.03
-50 -40 -30 -20 -10 0 10 20 30 40 50 60 70
Error Degrees C
Temperature Degrees C
Steinhart & Hart - Tabulated values