Section 7. Installation
160
1. Send the program in CRBasic example FieldCal Multiplier and Offset
Demonstration Program
(p. 160) to the CR800.
2. To simulate the flow sensor, place a jumper wire between channels VX1/EX1
and SE6 (3L).
3. Simulate deployment-calibration conditions (output @ 30 l/s = 300 mV, output
@ 10 l/s = 550 mV) in two stages.
a. Set variable SignalmV to 300. Set variable KnownFlow to 30.0.
b. Start the deployment calibration by setting variable CalMode = 1.
c. When CalMode increments to 3, set variable SignalmV to 550. Set
variable KnownFlow to 10.
d. Resume the deployment calibration by setting variable CalMode = 4
4. When variable CalMode increments to 6, the deployment calibration is
complete. Calibrated multiplier is -0.08. Calibrated offset is 53.978.
5. To continue this example, simulate a two-stage, 7-day service calibration
wherein both multiplier and offset drift (output @ 30 l/s = 285 mV, output @
10 l/s = 522 mV).
a. Set variable SignalmV to 285. Set variable KnownFlow to 30.0.
b. Start the 7-day, service calibration by setting variable CalMode = 1.
c. When CalMode increments to 3, set variable SignalmV to 522. Set
variable KnownFlow to 10.
d. Resume the 7-day service calibration by setting variable CalMode = 4
6. When variable CalMode increments to 6, the 7-day, service calibration is
complete. Calibrated multiplier is -0.0842. Calibrated offset is 53.896.
CRBasicExample29. FieldCal()MultiplierandOffsetDemonstrationProgram
'Jumper VX1/EX1 to SE6(3L) to simulate a sensor
Public SignalmV 'Excitation mV output
Public KnownFlow 'Known water flow
Public CalMode 'Calibration trigger
Public Multiplier 'Sensitivity
Public Offset 'Offset (starts at zero, not changed)
Public WaterFlow 'Water flow
'Data Storage Output of Calibration Data stored whenever a calibration occurs
DataTable(CalHist,NewFieldCal,200)
SampleFieldCal
EndTable
BeginProg
Multiplier = 1
Offset = 0
LoadFieldCal(true) 'Load the CAL File, if possible