Campbell-scientific 10164-L Water Sampler Control Cable for use with I Manual de usuario Pagina 10

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 18
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 9
10164-L Sampler Control Cable for use with Isco and Sigma Autosamplers
TABLE 3-4. Datalogger Wiring to Inhibit a
Sigma Sampler Program
Color
Description
CR800
CR850
CR1000
CR3000
CR500
CR510
CR10(X)
CR23X
21X
Purple Ground
G
Brown Control Port Control Port
(C1, C2,…)
Control Port
(C1, C2,…)
Control
Clear Shield
G
TABLE 3-5. Datalogger Wiring to Inhibit an
Isco Sampler Program
Color
Description
CR800
CR850
CR1000
CR3000
CR500
CR510
CR10(X)
CR23X
21X
Purple Ground
G
Green Control Port Control Port
(C1, C2,…)
Control Port
(C1, C2,…)
Control
Clear Shield
G
Insulate and tuck the unused wires out of the way.
4. Programming
The datalogger is programmed using either CRBasic or Edlog. Dataloggers
that use CRBasic include the CR800, CR850, CR1000, and CR3000.
Dataloggers that use Edlog include the CR10(X), CR510, CR500, CR23X, and
21X.
With this cable, the datalogger can be programmed to:
Trigger the sampler
Sense and record when the sampler has taken a sample
Inhibit the sampler from running its onboard sampling routine
Each of these functions is independent of the others and may be combined as
desired.