Appendix A. Analog Output Configurations
A-2
TABLE A-2. Dip Switches SW2: Analog Output Pressure Range Settings
RANGE SELECTION SWITCH SETTINGS
mbar mm Hg in. Hg
LOWER UPPER LOWER UPPER LOWER UPPER SW2-1 SW2-2 SW2-3 SW2-4
600 800 450 600 17.72 23.62 ON ON ON ON
600 900 450 675 17.72 26.58 OFF ON ON ON
600 1000 450 750 17.72 29.53 ON OFF ON ON
600
1
1100
1
450
1
825
1
17.72
1
32.48
1
OFF
1
OFF
1
ON
1
ON
1
700 800 525 600 20.67 23.62 ON ON OFF ON
700 900 525 675 20.67 26.58 OFF ON OFF ON
700 1000 525 750 20.67 29.53 ON OFF OFF ON
700 1100 525 825 20.67 32.48 OFF OFF OFF ON
677.1 1083.6 508.0 812.8 20.00 32.00 ON ON ON OFF
800 900 600 675 23.62 26.58 OFF ON ON OFF
800 1000 600 750 23.62 29.53 ON OFF ON OFF
800 1100 600 825 23.62 32.48 OFF OFF ON OFF
2 2
2
2
2
2
ON
2
ON
2
OFF
2
OFF
2
2 2
2
2
2
2
OFF
2
ON
2
OFF
2
OFF
2
900 1000 675 750 26.58 29.53 ON OFF OFF OFF
900 1100 675 825 26.58 32.48 OFF OFF OFF OFF
1
Factory configuration at Campbell Scientific
2
Settings with no pressure values are invalid settings. If the switches are set to either of these combinations,
the analog output will default to the full range of 600 to 1100 mbar.