Campbell-scientific 092 Barometric Pressure Sensor Manual de usuario Pagina 15

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 24
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 14
Model 092 Barometric Pressure Sensor
'Main Program
BeginProg
'Main Scan
Scan(5,Sec,1,0)
'Default Datalogger Battery Voltage measurement 'BattV'
Battery(BattV)
'Default Wiring Panel Temperature measurement 'PTemp_C'
PanelTemp(PTemp_C,_60Hz)
'092 Barometric Pressure Sensor measurement 'BP_mmHg'
VoltSE(BP_mmHg,1,mV5000,1,1,0,_60Hz,0.1,600)
BP_mmHg=BP_mmHg*0.75006
'Call Data Tables and Store Data
CallTable(Table1)
CallTable(Table2)
NextScan
EndProg
8. Maintenance
8.1 Cleaning
Periodic cleaning of the sintered filter at the base of the unit may be required in
dusty conditions. Remove the filter and clean with distilled water.
8.2 Calibration
Calibration is performed against a NIST traceable standard. Calibration
coefficients are stored in non-volatile memory at the factory. No user
calibration procedure is available. The sensor must be returned to Met One
Instruments when calibration is desired. Yearly calibrations are recommended.
For calibration, contact:
Met One Instruments
1600 Washington Blvd.
Grants Pass, OR 97526
Phone: (541) 471-7111
Fax: (541) 471-7116
Met One Instruments
3206 Main Street
Suite 106
Rowlett, TX 75088
Phone: (972) 412-4747
Fax: (972) 412-4716
9
Vista de pagina 14
1 2 ... 10 11 12 13 14 15 16 17 18 19 20 ... 23 24

Comentarios a estos manuales

Sin comentarios