CS125 Present Weather Sensor
26
Separate ID used as an extra identifier for a
particular CS125 on a network.
User alarm one activation state
0 = Alarm one disabled
1 = Alarm one enabled
0 = Check if distance is less than ‘User alarm 1
Distance’
1 = Check if distance is greater than ‘User alarm 1
Distance’
Distance value that alarm one will trigger against.
This value will correspond to metres or feet
depending upon which is selected in ‘Visibility
Unit’
User alarm two activation state
0 = Alarm two disabled
1 = Alarm two enabled
0 = Check if distance is less than ‘User alarm 2
Distance’
1 = Check if distance is greater than ‘User alarm 2
Distance’
Distance value that alarm one will trigger against.
This value will correspond to metres or feet
depending upon which is selected in ‘Visibility
Unit’
Baud rate for the main RS232/RS485 interface
0 = 115200 bps
1 = 57600 bps
2 = 38400 bps
3 = 19200 bps
4 = 9600 bps
5 = 2400 bps
6 = 1200 bps
Internal serial number for the CS125.
(Read only)
Unit the visibility value will be presented as.
M = metres
F = feet
Interval in seconds between outputs in continuous
mode. This value has no effect if polled mode has
been selected in ‘Measurement mode’
Selects polled or continuous modes.
In continuous mode the sensor will output a string
in the format as set by ‘Message Format’ at regular
intervals as defined by ‘Continuous Interval’.
0 = Continuous mode
1 = Polled mode
Level of detail outputted by polled or continuous
modes
0, 1 and 2 = Basic, partial or full visibility
messages
3, 4 and 5 = Basic, partial or full SYNOP messages
6, 7 and 8 = Basic, partial or full METAR
messages