PWC100 Present Weather
Sensor Calibrator
Revision: 3/12
Copyright © 2008-2012
Campbell Scientific, Inc.
Vista de pagina 0
1 2 3 4 5 6 ... 21 22

Indice de contenidos

Pagina 1 - Sensor Calibrator

PWC100 Present Weather Sensor CalibratorRevision: 3/12Copyright © 2008-2012Campbell Scientific, Inc.

Pagina 2

PWC100 Present Weather Sensor Calibrator 3.2 Starting the Tests All PWC100 tests can be performed without the need to access the PWS with a PC. Only

Pagina 3 - Warranty

PWC100 Present Weather Sensor Calibrator Step 2: Performing the visibility maximum limit check • Remove the PWC100 visibility calibration disk and

Pagina 4 - Assistance

PWC100 Present Weather Sensor Calibrator Step 3: Position 1: Center position Attaching the velocity calibration unit • Fix the PWC100 velocity calib

Pagina 5 - PWC100 Table of Contents

PWC100 Present Weather Sensor Calibrator Step 4: Position 2: Left • Move the calibrator into the left hand position by unscrewing the calibrator mo

Pagina 6

PWC100 Present Weather Sensor Calibrator Step 5: Position 3: Right • Move the calibrator into the right-hand position by unscrewing the calibrator

Pagina 7 - Calibrator

PWC100 Present Weather Sensor Calibrator Step 6: Final stage • Remember to reinstall the plastic blanking screws back into the PWS100 before leaving

Pagina 8 - 3. PWS Calibration

PWC100 Present Weather Sensor Calibrator Quick look LED status table Ready/complete state Running state Test state Left Center Right Left Center Ri

Pagina 9

PWC100 Present Weather Sensor Calibrator It is also recommended that the operator stands behind the laser hood while the PWS is active. 6. Electrical

Pagina 10 - 3.2 Starting the Tests

PWC100 Present Weather Sensor Calibrator 12

Pagina 11

Appendix A. Assembly Guidelines A.1 Assembly Guidelines The PWC100 visibility and velocity calibration disks can easily be removed for cleaning or r

Pagina 13

Appendix A. Assembly Guidelines FIGURE A-3. Attachment of the velocity calibrator A.2 Connector Pin Outs Connector Front View 4: Data 1: 0 V 2: +

Pagina 15

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Pagina 16 - 4. Cleaning Guide

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Pagina 17 - 6. Electrical Parameters

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Pagina 18

PWC100 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links t

Pagina 20 - A.2 Connector Pin Outs

PWC100 Present Weather Sensor Calibrator 1. Introduction The PWC100 is a customer field calibration validation device for the PWS100. The PWC100 cali

Pagina 21

PWC100 Present Weather Sensor Calibrator Care must be taken not to drop or scratch either the velocity or visibility units as this may cause errors th

Pagina 22

PWC100 Present Weather Sensor Calibrator 3.1 Installing and Connecting up the Calibrator • Remove the existing plastic blanking screws from their po