Campbell-scientific OBS500 Smart Turbidity Meter with ClearSensor Tech Manual de usuario Pagina 33

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 64
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 32
OBS500 Smart Turbidity Meter with ClearSensor™ Technology
universal response curve. However, there is residual nonlinearity that is
removed by calibration and by computation of a TU value with a 2nd-order
polynomial. This section explains how to do a turbidity calibration.
0 1000 2000 3000 4000
Tur bidity (TU)
0.0
0.2
0.4
0.6
0.8
1.0
Normalized OBS-3+ Response (OPV330 VCSEL)
FIGURE 7-5. Normalized response of OBS500 to AMCO Clear
®
turbidity. The inset shows the response function of a turbidity
sensor to high-sediment concentrations.
AMCO Clear
®
SDVB turbidity standards are used to calibrate an OBS500
sensor. SDVB standards are made for individual instruments. Standards made
for one model of turbidity meter cannot be used to calibrate a different model.
0 20000 40000 60000
SSC (mg/l)
0
1000
2000
3000
4000
Turbidity (NTU)
A
B C
23
Vista de pagina 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 63 64

Comentarios a estos manuales

Sin comentarios