NR-LITE Net Radiometer
2: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 60 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
3: Real Time (P77)
1: 110 Day,Hour/Minute (midnight = 0000)
4: Average (P71)
1: 1 Reps
2: 1 Loc [ Net_rad ]
5.3.4 CR10X Example Program with Wind Speed Correction
This example measures the NR-LITE using a single-ended input and
Instruction 1. The program also measures wind speed and applies the
correction factor as described in Section 1.4. Average net radiation and wind
speed are output every hour.
NR-LITE Wiring for Example 4
Color
Function
Example CR10X
Program Channels
Used
White Radiation Signal S.E. Channel 1
Green Signal Reference AG
Clear Shield G
;{CR10X}
;
*Table 1 Program
01: 60 Execution Interval (seconds)
;measure 05103 wind speed
1: Pulse (P3)
1: 1 Reps
2: 1 Pulse Channel 1
3: 21 Low Level AC, Output Hz
4: 1 Loc [ Wspd_m_s ]
5: .0980 Mult
6: 0 Offset
;measure NR-LITE net radiation
2: Volt (SE) (P1)
1: 1 Reps
2: 3** 25 mV Slow Range
3: 1* SE Channel
4: 2* Loc [ Rn_obs ]
5: 63.29 Mult ;Multiplier for specific sensor
6: 0 Offset ;in units of (W m
-2
)/mV (see above)
13