Appendix F. Installing the AC/DC
Power Adapter Kit
The AC/DC Power Adapter Kit is configurable within the CPEC200 system
enclosure to allow the CPEC200 to be powered from AC mains power. A
peripheral mounting kit (pn 16987) is necessary to install the AC/DC adapter
into the CPEC200 system enclosure. The mounting kit includes a bracket, a
Velcro
®
strap, and the necessary nuts and screws.
The following steps describe the mounting procedure.
1. Install the peripheral mounting kit inside the CPEC200 system
enclosure by threading the Velcro
®
strap through the bracket.
2. Place the bracket at the backplate of the enclosure below the
datalogger and attach it with the screws included in the mounting kit
(see FIGURE F-1).
FIGURE F-1. Peripheral mounting kit installed in CPEC200 enclosure
3. Place the power supply on the mounting bracket and secure the power
supply’s cable to the back of the mounting bracket with a wire tie as
shown in FIGURE F-2.
F-1