Appendix C. Output Variables
If there is some error condition that prevents the zero/span sequence from
starting, one of the following messages will be displayed:
ERROR: Cannot run Zero/Span sequence - CHECK_ZERO is False
ERROR: Cannot run Zero/Span sequence - Check valve temperature
ERROR: Cannot run Zero/Span - Check scrub module temperature
If the CPEC200 begins the startup process for a zero/span sequence, and then
detects an error, one of the following messages will be displayed:
Zero/Span startup manually aborted by turning pump on
ERROR: Zero/Span startup aborted - Check scrub module
temperature
ERROR: Zero/Span startup aborted - Check valve temperature
If the CPEC200 has already started the zero/span sequence and then detects
some error condition that causes the zero/span sequence to abort, one of the
following messages will be displayed:
Zero/Span sequence manually aborted by turning pump on
ERROR: Zero/Span sequence aborted - Check scrub module
temperature
ERROR: Zero/Span sequence aborted - Check valve temperature
If the CPEC200 attempts to set the IRGA zero or span as part of a zero/span
sequence and there is a problem identified (diag_cpec is nonzero), the
CPEC200 will skip the zero or span setting step, continue the rest of the
sequence, and display one of the following messages. If the user attempts to
set the zero or span manually this error or check is bypassed, and the zero or
span will be set regardless of the status of diag_cpec.
ERROR: Automatic Zero aborted - Check diagnostics
ERROR: Automatic CO2 Span aborted - Check diagnostics
Some of the messages relate to the use of the valve module in manual mode:
Valve operation not allowed when pump is on
EC mode restarted by turning pump on
ERROR: Valves disabled - Check valve temperature
ERROR: Valves disabled - Check press_offset
ERROR: Valves disabled - Check scrub module temperature
Some of the messages relate to the configuration of the EC155:
ERROR: Powering OFF EC155 gas head
Powering ON EC155 gas head
Configuring the EC100
Setting the zero
Setting the CO2 span
Setting the H2O span
The values stored in table message_log are shown in TABLE C-4. The value