Campbell-scientific 110PV Surface Temperature Probe Manual de usuario Pagina 9

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 32
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 8
110PV Surface Temperature Probe
2. Accuracy
The overall probe accuracy is a combination of the thermistor's
interchangeability specification and the accuracy of the bridge resistor. The
Steinhart-Hart equation used to calculate temperature has a negligible error
(Figure 2-1). In a "worst case" the errors add to an accuracy of ±0.2°C over
the range of -40° to 70°C; ±0.5°C over the range of 71°C to 105°C; and ±1°C
from 106°C to 135°C. The major error component is the interchangeability
specification (tolerance) of the thermistor. The bridge resistor has a 0.1%
tolerance with a 10 ppm temperature coefficient. Figures 2-2 to 2-7 show the
possible worst case probe and measurement errors.
FIGURE 2-1. Steinhart-Hart Error
3
Vista de pagina 8
1 2 3 4 5 6 7 8 9 10 11 12 13 14 ... 31 32

Comentarios a estos manuales

Sin comentarios