WindSonic Two-Dimensional Sonic Anemometer
To activate the SDI-12 Transparent Mode, on Control Port p, enter pX and
press the <enter> key. The datalogger will respond with “entering SDI-
12”. If any invalid SDI-12 command is issued, the datalogger will exit the
SDI-12 Transparent Mode.
To query the WindSonic4 for its current SDI-12 address, enter the
command ?!. The WindSonic4 will respond with the current SDI-12
address.
To change the SDI-12 address, enter the command aAb!; where a is the
current address from the above step and b is the new address. The
WindSonic4 will change its address and the datalogger will exit the SDI-
12 Transparent Mode.
To activate the SDI-12 Transparent Mode on Control Port p, enter pX and
press the <enter> key. Verify the new SDI-12 address by entering the “?!”
command. The WindSonic4 will respond with the new address.
To exit the SDI-12 Transparent Mode, enter .
FIGURE 7-1. SDI-12 Transparent Mode on an Edlog array-based
datalogger (CR10X) using control port 8 and changing the SDI-12
address from 0 to 1
23