  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - Present Weather Sensor

PWS100Present Weather SensorRevision: 3/12Copyright © 2006-2012Campbell Scientific, Inc.

Pagina 2

PWS100 Table of Contents vi C-3. PWS100 communication cable ... C-4 C-4. Enclosure wiring det

Pagina 3 - Warranty

Section 8. Functional Description 8-14

Pagina 4 - Assistance

Section 9. Maintenance 9.1 General The PWS100 Present Weather Sensor is a robust instrument that will provide years of uninterrupted weather monitor

Pagina 5 - PWS100 Table of Contents

Section 9. Maintenance 9-2 It is advisable to use airduster to blow any loose dust and dirt from the lenses as a first step. Using a lint free lens c

Pagina 6

Section 10. Troubleshooting 10.1 Introduction If your PWS100 seems to be operating incorrectly, there are a number of checks you can make to isolate

Pagina 7

Section 10. Troubleshooting 10.2.3 Ice has formed in the end of the hoods The heaters may not be working. Enter the command mode of the sensor, run

Pagina 8

Section 10. Troubleshooting checked that there are no spider’s webs on the instrument, especially in and around the detection volume and hoods of the

Pagina 9 - List of Figures

Section 10. Troubleshooting 10-4

Pagina 10 - List of Tables

Appendix A. PWS100 Output Codes The codes that are used in the PWS100 are taken from WMO Manual on Codes 1995 edition, Suppl. No. 5 (VIII.2005), Rec.

Pagina 11 - Section 1. Introduction

Appendix A. PWS100 Output Codes Clouds generally dissolving or becoming less developed during the past hour 01 - - - - - State of sky on the whole

Pagina 12

Appendix A. PWS100 Output Codes Blowing or drifting snow or sand, visibility less than 1 km 29 DRSA / BLSA / DRSN / BLSN - - - - FOG 30 FG - - - ≤1

Pagina 13 - 2.2 Laser Safety

Section 1. Introduction The PWS100 is a laser-based sensor that measures precipitation and visibility by accurately determining the size and velocity

Pagina 14

Appendix A. PWS100 Output Codes Freezing precipitation, slight or moderate2 47 -UP / UP P- / P P- / P i<5.34 - Freezing precipitation, slight

Pagina 15

Appendix A. PWS100 Output Codes Rain and snow, light 67 -SNRA P- SR- i<1.84 - Drizzle and snow, light 67 -SNDZ P- SL- i<0.64 - Rain, drizz

Pagina 16

Appendix A. PWS100 Output Codes Snow showers, moderate1 86 SHSN S S 1.0≤i<5.0 >0.5 mm Snow showers, heavy1 87 +SHSN S+ S+ i≥5.0 >0.5 mm

Pagina 17 - Section 4. Overview

Appendix A. PWS100 Output Codes Subjective Observations (Trappes, Paris, France, 14-16 May 1997) and the Working Group on Surface Measurements (Genev

Pagina 18

Appendix A. PWS100 Output Codes However it is also noted that non-mixed reports such as rain or snow may also contain a proportion of other types of

Pagina 19 - Section 5. Specifications

Appendix A. PWS100 Output Codes TABLE A-6. Intensity bounds for rain, drizzle and snow. Rain Drizzle Snow Rain, Drizzle and Snow Light i<2.5

Pagina 20 - 5.3 Optical Specifications

Appendix A. PWS100 Output Codes A-10

Pagina 21

Appendix B. Wiring EARTH GROUND FIGURE B-1. Underside of DSP enclosure TABLE B-1. Cable identifier Connector Number Connector Description Connect

Pagina 22

Appendix B. Wiring FIGURE B-2. DSP PCB to DSP enclosure connections B-2

Pagina 23 - Section 6. Installation

Appendix C. Cable Selection C.1 Power Cable The PWS100 is provided pre-wired with a default 10 m power cable as shown in Figure C-1. This cable shou

Pagina 24 - 6.2 Unloading and Unpacking

Section 1. Product Overview 1-2

Pagina 25 - 6.3 Installation Procedures

Appendix C. Cable Selection WIRING INSTRUCTIONS COLOR DSP SCREEN G BLACK G RED 12V POWER +24 Vdc – GREEN A/C OR DC -24 Vdc – WHITE A/C OR D

Pagina 26

Appendix C. Cable Selection C.2 Communication Cable The PWS100 has RS-232, RS-422 and RS-485 communications capability. Use a screened 2 × 0.22 mm2

Pagina 27

Appendix C. Cable Selection WIRING INSTRUCTIONS DSP COLOR 9WAY TX-Z BLUE 2 CTS-Y YELLOW 8 RTS-B WHITE 7 RX-A GREEN 3 G BLACK 5 G485 RED N/C G SCR

Pagina 28

Appendix D. Software Flowchart D-1

Pagina 29 - 6.3.4 Basic Wiring

Appendix D. Software Flowchart D-2

Pagina 30

Appendix D. Software Flowchart Filter FFT Channel 1Autocorrelation and DC removalFilter FFTChannel 2Autocorrelation and DC removalInverse FFT and sel

Pagina 31 - 6.3.6 Communication Options

Appendix D. Software Flowchart Filter FFTChannel 1Autocorrelation and DC removalInverse FFT and select real componentFind correlation peak positionSe

Pagina 32

Appendix D. Software Flowchart D-5

Pagina 33

Appendix D. Software Flowchart Determine equivolumetric diameter and volumeDetermine size and velocity table co-ordinatesDetermine fuzzy logic values

Pagina 34

Appendix E. Menu System Map E-1

Pagina 35 - 6.3.9 Initial Settings

Section 2. Cautionary Statements 2.1 Sensor Unit Safety The PWS100 sensor has been checked for safety before leaving the factory and contains no int

Pagina 36 - 6.4.3 Lightning Rod

Appendix E. Menu System Map E-2

Pagina 37 - Section 7. Operation

Appendix E. Menu System Map WELCOME TO THE PWS100 SETUP MENUID {id}SN {sn}(0) message 0(1) message 1(2) message 2(3) set time & date(4) configura

Pagina 38 - 7.3 Terminal Mode

Appendix E. Menu System Map E-4

Pagina 40 - 7.4 PWS100 Menu System

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Pagina 41

Section 2. Cautionary Statements If the laser is operated outside of the housing then the following warning applies: INVISIBLE LASER RADIATION AVOID

Pagina 42

Section 3. Initial Inspection Upon receipt of the PWS100, inspect the packaging and contents for damage. File damage claims with the shipping company

Pagina 43

Section 3. Initial Inspection 3-2

Pagina 44

Section 4. Overview The PWS100 Present Weather Sensor is a laser based sensor capable of determining precipitation and visibility parameters for auto

Pagina 45

Section 4. Overview 4-2

Pagina 46

Section 5. Specifications 5.1 Mechanical Specifications Measuring Area: 40 cm2 (6.2 in2) Housing Materials: Iridite NCP conversion coated alumin

Pagina 48

Section 5. Specifications 5.3 Optical Specifications 5.3.1 Laser Head Specifications Laser Source: Near-infrared (IR) diode, eye safe Class 1M un

Pagina 49

Section 5. Specifications 5.6.2 Precipitation Measurements Particle Size*: 0.1 mm to 30 mm (0.004 in to 1.18 in) Size Accuracy*: ± 5% (for parti

Pagina 50

Section 5. Specifications 5-4 The particle buffer is able to hold raw data for 500 typical particles. The processor is able to process the particles

Pagina 51

Section 6. Installation 6.1 Location and Orientation The PWS100 measures environmental variables and is designed to be located in harsh weather cond

Pagina 52

Section 6. Installation In order to minimize user interaction with the unit, the PWS100 should be placed away from sources of contamination, in the c

Pagina 53

Section 6. Installation 6.3 Installation Procedures 6.3.1 Assembling the PWS100 The PWS100 comes as a single unit, with the DSP enclosure attached

Pagina 54

Section 6. Installation Bracket TabU-bolt Bracket DSP Plate FIGURE 6-2. Hardware for mounting the top of the DSP plate to a pole 6-4

Pagina 55

Section 6. Installation Bracket Tab Notches FIGURE 6-3. Placing the PWS100 onto the bracket 6-5

Pagina 56

Section 6. Installation FIGURE 6-4. PWS100 mounted to a mast or pole Ensure that the PWS100 is mounted according to Figures 6-2 through 6-4. Do not

Pagina 57

Section 6. Installation 6.3.3 Connecting Cables The sensor unit comes with the DSP control unit fixed to the sensor arm. All cabling between the sen

Pagina 58

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Pagina 59

Section 6. Installation PG9 CABLE GLAND PG11 CABLE GLAND(HOOD HEATER) LEMO 4-PIN (CONNECTOR FORCS215-PWS) EARTH GROUNDFIGURE 6-5. Underside of DS

Pagina 60

Section 6. Installation 6.3.5 Desiccant The desiccant bags should be removed from the plastic bags in which they are shipped before placing them ins

Pagina 61

Section 6. Installation FIGURE 6-7. Removal of DSP cover. FIGURE 6-8. Exposing the DSP board. 6-10

Pagina 62

Section 6. Installation The location of the dip switches on the board is shown in Figure 6-9 and the dip switches themselves are shown in detail in F

Pagina 63

Section 6. Installation FIGURE 6-9. DSP board dip switch location (circled) FIGURE 6-10. Dip switches (defaults set - 00011100) 6-12

Pagina 64

Section 6. Installation 6.3.7 Installing Power Supply Power supply connections can be made in the PWS100 24 Vdc/12 Vdc power supply cabinet using th

Pagina 65

Section 6. Installation 6-14 replacing the desiccant. This is of particular importance if using the sensor in corrosive or salt laden atmospheres. 6.

Pagina 66

Section 7. Operation 7.1 Introduction The best way of becoming familiar with the sensor is to setup the sensor and connect it to a PC running Window

Pagina 67

Section 7. Operation configuration options. These are discussed below within the context of setting up the sensor using a terminal emulator. 7.3 Te

Pagina 68 - 0.178*D

Section 7. Operation In the descriptions which follow ↵ symbolizes the pressing of the ENTER key. Input parameters in italics should be user-defined

Pagina 69

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Pagina 70

Section 7. Operation The data storage function will not store data while the terminal or menu is active. This means data will be missing during these

Pagina 71

Section 7. Operation Section 7.4.6 gives further details of how to set up the PWS100 with the command set in the terminal mode (option 7 from the SET

Pagina 72

Section 7. Operation If option 1 is chosen from the MESSAGE menu then the system will display the MESSAGE PARAMETERS and FIELDS menu for that message

Pagina 73

Section 7. Operation From the MESSAGE PARAMETERS and FIELDS menu if field 1 is chosen then the MESSAGE MODE menu (see Figure 7-5) will be displayed.

Pagina 74

Section 7. Operation TABLE 7-2. Message Field parameters Message Field Parameter Output 34 External sensor (Aux) 40 Precipitation intensity (mmh-1

Pagina 75 - Wetness grid value

Section 7. Operation Note that user defined messages cannot make field references to other user defined messages but can make reference to fixed mess

Pagina 76

Section 7. Operation 7.4.1.3 Message Field 10 To 19 Fixed Messages No fixed messages have been defined yet. The user cannot change the fixed message

Pagina 77

Section 7. Operation 10. DC voltage of upper detector greater than 1.5 V. This can be caused by sun directly shining into lens or a pws fault. 11.

Pagina 78

Section 7. Operation The PWS100 only measures particles / visibility 90% of the time so precipitation intensity is scaled appropriately. NOTE 7.4.1.1

Pagina 79 - 7.8 Maintenance Commands

Section 7. Operation 15 =>4.50 0.50 16 =>5.00 0.50 17 =>5.50 0.50 18 =>6.00 0.50 19 =>6.50 0.50 20 =>7.00 93.00 Particle speed

Pagina 80

PWS100 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links t

Pagina 81 - 7.9 Other Commands

Section 7. Operation 7 0.795 0.13 8 0.925 0.13 9 1.055 0.13 10 1.185 0.13 11 1.375 0.25 12 1.625 0.25 13 1.875 0.25 14 2.125 0.25 15 2.375 0.25 16 2.

Pagina 82 - 7.10.1 Connections

Section 7. Operation 10 0.95 0.1 11 1.10 0.2 12 1.30 0.2 13 1.50 0.2 14 1.70 0.2 15 1.90 0.2 16 2.20 0.4 17 2.60 0.4 18 3.00 0.4 19 3.40 0.4 20 3.80

Pagina 83

Section 7. Operation 8 =>0.70 0.1 9 =>0.80 0.1 10 =>0.90 0.1 11 =>1.00 0.2 12 =>1.20 0.2 13 =>1.40 0.2 14 =>1.60 0.2 15 =>1.8

Pagina 84

Section 7. Operation 11 =>1.00 0.2 12 =>1.20 0.2 13 =>1.40 0.2 14 =>1.60 0.2 15 =>1.80 0.2 16 =>2.00 0.4 17 =>2.40 0.4 18 =>2

Pagina 85

Section 7. Operation 7.4.1.24 Message Field 100 Upper, Lower LED temperature This field will output the sampled internal upper and lower LED tempera

Pagina 86

Section 7. Operation 7.4.1.35 Message Field 154 Watchdog Count, Maximum Particles Per Second, Particles Not Processed, Time Lag This field will outp

Pagina 87 - 8.2 Optical Measurement

Section 7. Operation FIGURE 7-6. Message field menu Choosing option 0 on the MESSAGE FIELD menu will display more output parameter options. Choosin

Pagina 88 - Doublet lens Grating Laser

Section 7. Operation FIGURE 7-7. Delete message menu 7.4.2 Top Menu Option 3 (Set Time and Date) Choosing option 3 from the SETUP menu brings up t

Pagina 89

Section 7. Operation 7.4.3 Top Menu Option 4 (Configuration) Choosing option 4 from the SETUP menu brings up the CONFIGURATION menu (see Figure 7-9)

Pagina 90

Section 7. Operation FIGURE 7-10. PWS100 ID menu Option 2 of the configuration menu gives the TRH PROBE TYPE menu (see Figure 7-11). Choose the cor

Pagina 91 - 8.4 PWS100 Control Unit

PWS100 Table of Contents 6.4 Grounding and Lightning Protection ... 6-14 6.4.1 Equipment Grounding...

Pagina 92 - 8.6 Algorithm Description

Section 7. Operation Option 3 of the configuration menu gives the WETNESS PROBE TYPE menu (see Figure 7-12). Choose the correct wetness probe that is

Pagina 93 - Dr izzle

Section 7. Operation Option 5 of the configuration menu gives the HOOD HEATER TEMPERATURE menu (see Figure 7-14). The value that is to be set needs t

Pagina 94

Section 7. Operation Option 7 of the configuration menu gives the OUTPUT MODE menu (Figure 7-16). This refers to the use of packetized output. The me

Pagina 95

Section 7. Operation FIGURE 7-17. Calibration warning screen Once confirm is typed followed by the return key, the CALIBRATION top menu (see Figure

Pagina 96 - 8.6.4 Present Weather

Section 7. Operation FIGURE 7-19. Calibration disc constants menu Selecting option 1 on the CALIBRATION top menu will bring up the VIEW / ADJUST CA

Pagina 97 - 8.6.5 Visibility

Section 7. Operation FIGURE 7-20. View / adjust calibration menu Option 9 of the configuration menu gives the TERMINAL MODE menu (see Figure 7-21).

Pagina 98 - 8.7 Applications

Section 7. Operation Option 10 of the configuration menu gives the PSU SHUT DOWN VOLTAGE menu (Figure 7-22). Enter the PSU input voltage level below

Pagina 99 - 8.8 Internal Monitoring

Section 7. Operation 7.4.5 Top Menu Option 6 (Weather and Alarm Parameters) Choosing option 6 from the SETUP menu brings up the WEATHER AND ALARM PA

Pagina 100

Section 7. Operation The snow water content scaling factor (SWCF) can be adjusted by choosing option 4 on the WEATHER PARAMETERS menu. This will brin

Pagina 101 - Section 9. Maintenance

Section 7. Operation FIGURE 7-26. Snow water content adjustment The mixed precipitation threshold value can be adjusted by choosing options 5-8 on

Pagina 102 - 9.3 Calibration

PWS100 Table of Contents 7.4.1.29 Message Field 105 DSP PSU Voltage, Hood and Dew Heater % Duty ...

Pagina 103 - Section 10. Troubleshooting

Section 7. Operation FIGURE 7-27. Mixed precipitation threshold adjustment 7.4.6 Top Menu Option 7 (Terminal) Choosing option 7 from the SETUP men

Pagina 104

Section 7. Operation of the operating system or a hardware reset. The SDI-12 values of temperature and relative humidity will be -999.00 if no such i

Pagina 105

Section 7. Operation FIGURE 7-30. Done menu 7.5 Message Related Commands Messages can be set for a variety of output types, including fixed standa

Pagina 106

Section 7. Operation The Message_Interval parameter must also be defined from 0 second to 32767 seconds which is the rate at which the system will di

Pagina 107

Section 7. Operation To set manual polling mode the MSET command should be of the form: MSET Message_ID 0 0 Message_Fields ↵ e.g., MSET 0 0 0 105

Pagina 108

Section 7. Operation Higher codes indicate other type of data (stored every 10 seconds) where values a, b, and c have different meanings as indicated

Pagina 109

Section 7. Operation 7.5.3 Viewing Data Output on the Command Line A number of records (defined by the number n) of uncollected historic m data (mes

Pagina 110

Section 7. Operation To set the adjustable visibility limit parameters, type: SETPARAM vislim1 vislim2 vislim3 snowwater mixthreshold↵ The system w

Pagina 111

Section 7. Operation The PWS100 will then display the configuration parameters on screen in the following format: Date and Time PWS100 product identi

Pagina 112

Section 7. Operation Currently three other sensors can be connected to the PWS100. The sensor configured is changed by parameters TRH_Sensor (Tempera

Pagina 113

PWS100 Table of Contents 8.3 Additional Sensor Connections... 8-3 8.3.1 Using a CS215-PWS

Pagina 114

Section 7. Operation Select the Xmodem 1k protocol and select the OS file using the ‘Browse’ button. Once selected press the ‘Send’ button. The Xmode

Pagina 115

Section 7. Operation 7.8.3 Running the Calibration The detection volume is calibrated using the PWC100 calibrator. The calibration fixture plate whi

Pagina 116

Section 7. Operation To set the time without the date being altered the following should be used: TIME hh:mm:ss↵ 7.9.2 Resetting the System The sys

Pagina 117 - Appendix B. Wiring

Section 7. Operation relative gains or losses of the two clocks that the logger will over time either receive an extra transmission from the sensor o

Pagina 118

Section 7. Operation 'Main Program BeginProg 'Open Serial Port to the PWS100 SerialOpen (Com1,115200,3,0,10000) 'Commands can b

Pagina 119 - Appendix C. Cable Selection

Section 7. Operation 'Open Serial Port to the PWS100 SerialOpen (Com1,115200,3,0,10000) 'Commands can be included here to setup the s

Pagina 120 - WIRING INSTRUCTIONS

Section 7. Operation 7-50

Pagina 121 - C.2 Communication Cable

Section 8. Functional Description 8.1 General The PWS100 Present Weather Sensor is an optical sensor using the best of scatter meter and disdrometer

Pagina 122

Section 8. Functional Description FIGURE 8-1. Laser unit Rod lens Cylindrical lens Doublet lens Grating Laser FIGURE 8-2. Laser unit showing l

Pagina 123

Section 8. Functional Description FIGURE 8-3. Sensor unit Plano-convex lens Photodiode Filter FIGURE 8-4. Sensor unit showing light path extents

Pagina 124

PWS100 Table of Contents List of Figures 4-1. PWS100... 4

Pagina 125

Section 8. Functional Description 8.3.1 Using a CS215-PWS on the PWS100 The CS215-PWS temperature and RH probe should be connected directly to one o

Pagina 126 - Polycrystalline processing

Section 8. Functional Description 8.4 PWS100 Control Unit The PWS100 control unit, shown in Figure 8-5 as a block diagram, is a custom designed DSP

Pagina 127

Section 8. Functional Description 8.6 Algorithm Description 8.6.1 Detecting and Classifying Precipitation The PWS100 has a structured detection vol

Pagina 128

Section 8. Functional Description Signal to Pedestal Ratio Analysis01002003004005006001 1.5 2 2.5 3 3.5 4 4.5 5Signal to Pedestal RatioParticle Count

Pagina 129 - Appendix E. Menu System Map

Section 8. Functional Description size/velocity value of 0 for drizzle, 0.4 for rain, 0.3 for snow flakes, 0.4 for graupel, 0.1 for ice pellet etc. T

Pagina 130

Section 8. Functional Description The PWS100 has exceptional sensitivity to the start and end of precipitation events. To avoid falsely reporting pr

Pagina 131

Section 8. Functional Description 8.6.3 Precipitation Accumulation Precipitation accumulation is calculated in millimeters over a specified time per

Pagina 132

Section 8. Functional Description 8.6.4.2 Visibility Types Also by using the WMO SYNOP code table (4680) a visibility type can be defined. These typ

Pagina 133

Section 8. Functional Description possible to measure the immediate surroundings, selecting appropriate parameters, which can be related to the envir

Pagina 134

Section 8. Functional Description 8.8 Internal Monitoring The PWS100 has a number of internal checks including temperature analysis inside each head