CS510 OxyGuard Dissolved Oxygen Probe
3
3. Service Kit
Spare membrane cap with fitted membrane.
50 ml electrolyte.
Spare anode and O-ring.
FIGURE 3-1. Exploded View of CS510-L
4. Optional Probe Accessories
Agitator for stagnant conditions.
Probe Mounting assembly for waste water installations.
Flow-through and by-pass fittings.
EasyCal calibrator for fast and accurate calibration.
5. Optional Agitator
The PT4 Agitator is a reliable and robust agitator for use in conjunction with
probes subjected to bio-fouling in ponds and stagnant water conditions.
O
2
probes require a minimum water velocity across their membranes to function
properly. Therefore, to measure DO in stagnant water conditions, it is necessary to
move the water past the membrane to get accurate and reliable DO measurements.
In many instances the water also has a high bio-loading and the probes become
fouled resulting in inaccurate DO measurements.