CS470/CS471 Compact Bubbler System
7.1 DIP Switch Settings
The default DIP switch settings are shown in TABLE 7-1. The default settings
support SDI-12 communications, which is the communication method typically
used by our dataloggers. The DIP switch settings need to be changed if using 4
to 20 mA communications (see Appendix C, 4 to 20 mA Operation).
TABLE 7-1. Dip Switch Settings for
SDI-12 Communications
SDI-12 Settings Switch Settings
1 ON
2 ON
3 OFF
4 NA
5 NA
6 NA
7 NA
8 NA
7.2 Installation/Water Depth Considerations
The CS470/CS471 and bubble chamber should be installed so that the tubing
has a continuous drop from the CS470/CS471 towards the bubble chamber.
Otherwise, moisture can collect in a hallow and potentially block the tubing
with the formation of drops.
The bubble chamber must be installed below the water at a fixed depth. The
maximum water depth is determined by the maximum tube length, which is
15.24 m (50 ft).
7