AVW1/AVW4
5-3
02: P55 Polynomial
01: 4 Reps
02: 1 X Loc TEMP C #1
03: 1 F(X) Loc [:TEMP C #1]
04: -104.78 C0
05: 378.11 C1
06: -611.59 C2
07: 544.27 C3
08: -240.91 C4
09: 43.089 C5
03: P28 Vibrating Wire (SE)
01: 4 Reps
02: 5 IN Chan
03: 1 Excite all reps w/EXchan 1
04: 24 Starting Freq. (units=100 Hz)
05: 31 End Freq. (units=100 Hz)
06: 500 No. of Cycles
07: 500 Rep delay (units=.01sec) 5
SEC DELAY BETWEEN
MEASUREMENTS
08: 5 Loc [:P psi #1]
09: 1 Mult
10: 0 Offset
04: P53 Scaling Array (A*loc +B)
APPLY MULT & OFFSET
01: 5 Start Loc [:P psi #1]
02: -15.1 A1
03: 142.4 B1
04: -15.1 A2
05: 142.4 B2
06: -15.1 A3
07: 142.4 B3
08: -15.1 A4
09: 142.4 B4
05: P53 Scaling Array (A*loc +B)
LOADS TEMP. COEFF.
01: 13 Start Loc [:Cpsi/C #1]
02: 0 A1
03: -.0698 B1
04: 0 A2
05: -.0698 B2
06: 0 A3
07: -.0698 B3
08: 0 A4
09: -.0698 B4
06: P87 Beginning of Loop
TEMPERATURE
CORRECTION
01: 00 Delay
02: 4 Loop Count
07: P34 Z=X+F ASSUMES
CALIBRATION TEMP. = 24
o
C
01: 1-- X Loc TEMP C #1
02: -24 F
03: 17-- Z Loc [:T-To*C #1]
08: P36 Z=X*Y
01: 13-- X Loc Cpsi/C #1
02: 17-- Y Loc T-To*C #1
03: 17-- Z Loc [:T-To*C #1]
09: P33 Z=X+Y
01: 5-- X Loc psi #1
02: 17-- Y Loc T-To*C #1
03: 9-- Z Loc [:Pt psi #1]
10: P95 End
FOLLOWING THREE INSTRUCTIONS
SUBTRACT THE BAROMETRIC PRESSURE
FROM THE THREE WATER PRESSURES
11: P35 Z=X-Y
01: 10 X Loc Ptb psi #2
02: 9 Y Loc Pt psi #1
03: 10 Z Loc [:Ptb psi #2]
12: P35 Z=X-Y
01: 11 X Loc Ptb psi #3
02: 9 Y Loc Pt psi #1
03: 11 Z Loc [:Ptb psi #3]
13: P35 Z=X-Y
01: 12 X Loc Ptb psi #4
02: 9 Y Loc Pt psi #1
03: 12 Z Loc [:Ptb psi #4]
First 99 Input Location Labels:
1:TEMP C #1 11:Ptbpsi #3
2:TEMP C #2 12:Ptbpsi #4
3:TEMP C #3 13:Cpsi/C #1
4:TEMP C #4 14:Cpsi/C #2
5:P psi #1 15:Cpsi/C #3
6:P psi #2 16:Cpsi/C #4
7:P psi #3 17:T-To*C #1
8:P psi #4 18:T-To*C #2
9:Pt psi #1 19:T-To*C #3
10:Ptbpsi #2 20:T-To*C #42