Campbell-scientific CS650 and CS655 Water Content Reflectometers Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 56
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
CS650 and CS655 Water Content Reflectometers
FIGURE 6-1. CS650G Insertion Guide Tool
6.3 Wiring
CS650 connections to a datalogger are shown below. Dataloggers are divided
into those which are programmed with the CRBasic programming language
and those that are programmed with Edlog. CRBasic dataloggers include the
CR1000, CR3000, CR5000, CR800-series, and CR200X-series. Compatible
Edlog dataloggers include the CR10X, CR23X, and CR510.
6.3.1 SDI-12 Wiring
TABLE 6-1 shows the SDI-12 wiring code for the CS650 water content
reflectometer. SDI-12 data is transmitted to a CRBasic datalogger odd
numbered control port or to any control port of an Edlog datalogger that is
capable of SDI-12 communication. See Section 6.6, Program Examples, for
SDI-12 programming examples.
TABLE 6-1. CS650 Wiring Code for SDI-12
Color Function Datalogger Connection
Green SDI-12 Data SDI-12 Input or Control Port
Red SDI-12 Power 12 Vdc
Black SDI-12 Reference G
Clear Shield G
Orange Not Used G
8