CS547 Conductivity and Temperature Probe and A547 Interface
18
02: Polynomial (P55)
01: 1 Rep
02: 11* X Loc Temp_C
03: 11* F(X) Loc [:Temp_C ]
04: -53.46 C0
05: 90.807 C1
06: -83.257 C2
07: 52.283 C3
08: -16.723 C4
09: 2.211 C5
* Proper entries will vary with program and datalogger channel and input location
assignments.
** On the 21X and CR7 use the 15 mV input range and 4000 mV excitation.
*** Use a multiplier of 0.2 with a 21X and CR7.
13. CS547 Schematic
Black (Ex Cond)
Green (Ex Temp)
Red (Temp)
Orange (Cond)
FIGURE 13-1. CS547 Conductivity and Temperature Circuit Diagram